Fluke
Fluke 750P24 Differential Pressure Module 0-15 psi measurement tool
Measures differential pressure from 0 to 15 psi with a 4X burst rating and 316 stainless steel hi-side media for use with Fluke 725, 726 and 750 series calibrators
- TypeElectronic
- Turns compatible calibrators into pressure calibrators
- 15 psi measurement range with 4X burst rating
- 316 stainless steel hi-side media compatibility
- Swappable between calibrators without accuracy loss
- 11 engineering units selectable at the touch of a button
Application and compatibility
The Fluke 750P24 differential pressure module converts compatible Fluke calibrators into pressure calibrators. It suits technicians who need to measure low-pressure ranges in industrial or laboratory settings. The module connects without tools and displays readings in up to 11 engineering units. It is not for users who require a standalone pressure gauge without an existing Fluke calibrator.
Pressure range and construction
The module measures differential pressure from 0 to 15 psi, equivalent to 0 to 1 bar or 0 to 100 kPa. Its high-side wetted parts are 316 stainless steel while the low side accepts dry air. A burst rating of 4X provides a safety margin above the maximum range. No power consumption or noise figures are provided for this passive sensor module.
Deciding specification
The critical figure is the 0 to 15 psi differential range with 316 stainless steel high-side compatibility. Buyers must confirm their process media suits these materials and that the pressure span covers the application. The module requires a Fluke 725, 726 or 750 series calibrator to operate. It will not function as an independent measuring device.
Highlights
- Turns compatible calibrators into pressure calibrators
- 15 psi measurement range with 4X burst rating
- 316 stainless steel hi-side media compatibility
- Swappable between calibrators without accuracy loss
- 11 engineering units selectable at the touch of a button
Specifications
| Brand | FLUKE |
|---|---|
| Type | Electronic |
Questions about this item
How do I install the 750P24 module on my 725 calibrator without tools?
Connect the module to a pressure source or hand pump, plug the module cable into the calibrator, then apply pressure; the no-tools connectors snap on without a wrench.
What maintenance will the 316 SS hi-side media need over time?
The 316 stainless steel hi-side port resists corrosion; periodic cleaning and seal inspection keep it accurate, and the independently calibrated module can be swapped between calibrators without recalibration.
At what pressure does the 4X burst rating become a limit during sustained use?
The module is rated to withstand four times the 15 psi range, so sustained pressure beyond that multiple risks permanent damage.