We take cryptocurrency — so the coins you hold can buy real hardware.Pay in crypto, spend it on real hardware. Free insured shipping worldwide over $399. Plain, unbranded boxes, sent from Germany.Free insured shipping over $399. Your rate is locked for 30 minutes once checkout opens.Rate locked 30 minutes. Every item sealed, serial-checked and covered by a 24-month warranty.Sealed, 24-month warranty.

Saved items Sign in

HNZXIB

HNZXIB 100ml Hydrothermal Synthesis Reactor 240°C PTFE lining KH series

SKU CH-96C0-C7FTP

A 100 ml hydrothermal synthesis reactor rated for 240 °C with a PTFE lining and non-magnetic stainless steel body

  • TypePressure Reactor
  • Temperature RangeRT~240°C
  • MaterialSteel
  • Capacity100ml
  • 100 ml capacity handles typical lab sample volumes
  • PTFE inner lining resists both acid and alkali attack
  • Non-magnetic 0Cr18Ni9Ti stainless steel shell avoids field interference
  • Operating limit of 240 °C with 220 °C safety ceiling
  • Heating and cooling rates kept below 5 °C/min for controlled runs

100 ml Hydrothermal Synthesis Reactor

This 100 ml hydrothermal synthesis reactor provides a sealed high-pressure vessel for digesting insoluble materials in acid or alkali environments. It suits university teaching labs, research units and small-scale experiments in new materials, energy or environmental engineering. The KH-series autoclave uses a non-magnetic stainless steel shell with a PTFE inner lining for double corrosion protection.

Temperature and Pressure Limits

The reactor operates from room temperature up to 240 °C with a security temperature of 220 °C. Heating and cooling rates must stay below 5 °C per minute to avoid damage. Safety pressure is rated at 3 Mpa and the filling degree must not exceed 80 % of the 100 ml capacity. Natural cooling to room temperature is required before opening the lid.

PTFE Lining Chemical Resistance

The Teflon inner liner resists both acid and alkali attack during hydrothermal digestion of heavy metals, pesticide residues, food, sludge, rare-earth products and organic matter. This chemical compatibility decides whether the vessel suits your specific sample matrix. Buyers needing larger volumes or different liner materials should look elsewhere.

Highlights

  • 100 ml capacity handles typical lab sample volumes
  • PTFE inner lining resists both acid and alkali attack
  • Non-magnetic 0Cr18Ni9Ti stainless steel shell avoids field interference
  • Operating limit of 240 °C with 220 °C safety ceiling
  • Heating and cooling rates kept below 5 °C/min for controlled runs

Specifications

TypePressure Reactor
Temperature RangeRT~240°C
MaterialSteel
Capacity100ml

Questions about this item

Should I choose the 100 ml capacity or a different volume?

The reactor is supplied as a fixed 100 ml vessel; the supplier notes that capacity volume is customizable, so other sizes must be ordered as a special request.

What does the 240 °C maximum temperature mean for daily use?

It is the absolute ceiling — heating and cooling rates must stay below 5 °C/min, and the safety temperature is 220 °C, so routine runs should be planned below that limit.

How does the PTFE lining affect chemical compatibility and pressure limits?

The Teflon inner liner resists acids and alkalis, while the stainless steel shell handles the pressure; safety pressure is rated at 3 Mpa and filling must not exceed 80 % of the 100 ml volume.